Automated indirect immunofluorescence antinuclear antibody analysis is a standardized alternative for visual microscope interpretation

Clin Chem Lab Med. 2013 Sep;51(9):1771-9. doi: 10.1515/cclm-2013-0016.

Abstract

Background: Screening for antinuclear antibodies (ANA) is a basic tool in the serological work-up of systemic rheumatic disorders. Despite the emergence of alternative screening methods and the difficulties in standardization, indirect immunofluorescence (IIF) remains the recommended method for ANA detection. This study aimed to assess the reliability of automated ANA IIF analysis as a standardized alternative for the conventional manual approach.

Methods: ANA testing on HEp-2000 cells was performed on 304 consecutive routine sera, 28 serumbank samples displaying rare staining patterns, 219 samples of well-defined disease cohorts [141 systemic sclerosis (SSc), 13 polymyalgia rheumatica, 22 osteoarthritis, 5 ANCA-associated vasculitis and 38 spondyloarthritis] and 100 healthy donors. All samples were analyzed by automated IIF (Zenit G-sight), by conventional visual IIF microscopy and two ANA screening enzyme immunoassays (EIA).

Results: Automated and conventional ANA IIF analysis were comparable for negative/positive interpretation as well as intensity assessment (>90% agreement). In contrast, the accuracy of pattern recognition (26%) was limited. Likelihood ratios (LR) for SSc on results intervals of both Zenit G-sight and EIA increased with increasing level of positivity. Sensitivity within the SSc-associated antibody subsets was higher for Zenit G-sight (97%-100%) than EIA (10%-96%). A significant correlation between the quantitative result obtained by Zenit G-sight and the conventional end-point titer was found.

Conclusions: The use of Zenit G-sight for automated ANA IIF analysis offers opportunities to improve standardization. However, a complementary role of the expert technicians remains, especially for pattern recognition and classification of uncertain/negative samples.

Publication types

  • Comparative Study

MeSH terms

  • Antibodies, Antinuclear / blood*
  • Automation
  • Cohort Studies
  • Fluorescent Antibody Technique, Indirect / methods*
  • Humans
  • Immunoenzyme Techniques
  • Microscopy / methods
  • Reference Standards

Substances

  • Antibodies, Antinuclear